Timing Chart
FAIL STOP mode timing chart
When the test duration setting (TIMER) is ON and the measurement is performed by the /START
signal input from EXT.I/O or when the test duration setting (TIMER) is ON and the measurement
is performed by the START key on the panel
Setting
Test duration (TIMER) Response time (DELAY) TEST signal OFF timing
ON, 45 ms to 999.9 s AUTO, 5 ms to 999.9 s FAST
When the test duration (TIMER) is set to OFF, tests are performed continuously
until a FAIL judgment is made.
Measurement-1 Measurement-2
10 V
t1
t2
t6
t3 t5
t4
/START
(Input)
/TEST
(Output)
Test voltage
Measured
value display
/PASS
(Output)
/U.FAIL
/L.FAIL
(Output)
/ERR
(Output)
t7
Measurement end
When the TEST signal OFF
timing setting is SLOW
When measured
value-1 result is PASS
When measured
value-2 result is FAIL
Previous measured value
Previous result
Previous result
Previous result
OFF ON OFF
OFF OFF ON
OFF
0/0.0/0.00 M
Ω
Measured value-1 Measured value-2
Measurement-1 contact check result
Measurement,
judgment
(including
contact check)
10
9
8
7
6
5
4
3
2
1