94
Timing Chart
Item Time
t1 START, STOP signal pulse width 5 ms
MIN.
t2 START, STOP signal detection time 5 ms
MAX.
*
t3 Response time (DELAY) AUTO, 5 ms to 999.9 s
t4 Measurement
time
Contact check: OFF 30 ms (FAST), 480 ms (SLOW)
Contact check: ON 80 ms (FAST), 480 ms (SLOW)
t5 Measurement
interval
Contact check: OFF 50 ms (FAST), 500 ms (SLOW)
Contact check: ON 100 ms (FAST), 500 ms (SLOW)
t6 Discharge time (time for the output voltage to drop
below 10 V)
20 ms
MAX.
(during pure resistance
measurement)
t7 Test duration (TIMER) 45 ms to 999.9 s
*
When the START signal is input after the test voltage is changed, a time of up to 500 ms is
added to detect the START signal.
Forced termination judgment mode timing chart
When the test duration setting (TIMER) is OFF and the measurement is performed by the /START
signal or /STOP signal input from EXT.I/O or when the test duration setting (TIMER) is OFF and
the measurement is performed by the START key or STOP key on the panel
Setting
Test duration (TIMER) Response time (DELAY) TEST signal OFF timing
OFF AUTO, 5 ms to 999.9 s FAST