131
Command Reference
:STATe?
Returns the instrument status in numerical data (<data>). No command header is included in
response.
Syntax
:STATe?
Response
<data>
<data> 0 to 2 (NR1 numerical data)
0: Stopped, 1: Testing, 2: Discharging
Example To query the status (during tests)
PC>
:STATe?
ST5520>
1
:MEASure?
Queries the measured value. When this command is received during a test, the measured value
is returned in NR3 numerical data (<data>).
After the test is completed, the measured value when the previous test was completed is
returned. No command header is included in response.
Syntax
:MEASure?
Response
<data>
<data> Measured value “8.5 Data Format Table” (p. 125)
Example To query the measured value (when the measured value is 123.4 M
Ω
)
PC>
:MEASure?
ST5520>
123.4E+06
10
9
8
7
6
5
4
3
2
1
Appx. Ind.