EasyManua.ls Logo

Hioki ST5520-01 - 9 Specifications; Environment and Safety Specifications

Hioki ST5520-01
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
143
Command Reference
:SYSTem:LFRequency
Sets the power frequency.
When AUTO is set, 50 Hz and 60 Hz are automatically identied.
Syntax
:SYSTem:LFRequency <data>
<data> AUTO/50/60 (text data)
AUTO: The power frequency is automatically identied.
50: Power frequency 50 Hz
60: Power frequency 60 Hz
Example To set the power frequency to 60 Hz
PC>
:SYSTem:LFRequency 60
:SYSTem:LFRequency?
Queries the power frequency setting. The setting is returned in NR1 number data (<data>).
Syntax
:SYSTem:LFRequency?
Response
:SYSTEM:LFREQUENCY <data>
or simply
<data>
<data> AUTO/50/60 (text data)
Example To query the power frequency setting (when the power frequency is 60 Hz)
PC>
:SYSTem:LFRequency?
ST5520>
:SYSTEM:LFREQUENCY 60
(when the
:HEADer
command is ON)
ST5520>
60
(when the
:HEADer
command is OFF)
:PANel:LOAD
Loads the test conditions with the specied number. During a test, the test is stopped and the
test conditions are loaded.
Syntax
:PANel:LOAD <data>
<data> Test condition number 1 to 10 (NR1 numerical data)
Example To load No. 3 test conditions
PC>
:PANel:LOAD 3
:PANel:SAVE
Saves the test conditions with the specied number.
Syntax
:PANel:SAVE <data>
<data> 1 to 10 (NR1 numerical data)
Example To save as No. 3
PC>
:PANel:SAVE 3
10
9
8
7
6
5
4
3
2
1
Appx. Ind.

Table of Contents

Related product manuals