Command Reference
:Panel:CLEar
Deletes the test conditions.
Syntax
:Panel:CLEar <data1>
<data1> Test condition number 1 to 10 (NR1 numerical data)
Example To delete Test condition 1
PC>
:Panel:CLEar 1
:AOUT:RANGe
Sets whether the analog output is to be output in all measurement resistance ranges or in each
range display area.
Syntax
:AOUT:RANGe <data>
<data> FULL/EACH (text data)
FULL: Output in all measurement resistance ranges
EACH: Output in each range display area
Example To set to output in all measurement resistance ranges
PC>
AOUT:RANGe FULL
:AOUT:RANGe?
Queries the analog output setting.
The FULL or EACH setting is returned in text data (<data>).
Syntax
:AOUT:RANGe?
Response
:AOUT:RANGE <data>
or simply
<data>
<data> FULL/EACH (text data)
FULL: Output in all measurement resistance ranges
EACH: Output in each range display area
Example To query the output method (when the setting is FULL)
PC>
:AOUT:RANGe?
ST5520>
:AOUT:RANGE FULL
(when the
:HEADer
command is ON)
ST5520>
FULL
(when the
:HEADer
command is OFF)
:PROBe
Sets the 9299 Switched Probe operation mode.
During a test, the test is stopped and the operation mode is set.
Syntax
:PROBe <data>
<data> CONTinue/TRIGger (text data)
CONTinue: Continuous mode
TRIGger: Trigger mode
Example To set continuous mode
PC>
: PROBe CONTinue