31
Pre-Operation Inspection
Checking LOW side contact
Perform the following check with the LOW terminal test leads removed.
1
Turn off the instrument power.
2
Connect the test leads to the HIGH
terminals.
(Do not connect the test leads to the LOW
terminals.)
LOW
Red
HIGH
For contact
check
3
Turn on the instrument power.
4
Set the contact check function to [ON].
(p. 56)
5
Allow the test leads to short circuit.
6
Start a test.
7
Check that a contact error occurs.
(If the test still starts even under these conditions,
the contact check function may not be [ON] or the
instrument may be damaged.)