32
Pre-Operation Inspection
Checking the operation of the contact check function
1
Turn off the instrument power.
2
Connect the test leads to the LOW and
HIGH terminals.
LOW
Black
HIGH
Red
For contact check
3
Turn on the instrument power.
4
Set the contact check function to [ON].
(p. 56)
5
Allow the test leads to short circuit.
(Short circuit the black test leads and red test
leads. Do not short circuit the black test lead and
red test lead.)
6
Start a test.
7
Check that the test starts.
(If the test does not start, the test leads may be
broken or the instrument may be damaged.)