63
Checking Short Circuit before Applying the Set Voltage (Short circuit check function)
The short circuit check function is operated using 100 k
Ω
as a criterion.
Therefore, if equipment operating at 100 k
Ω
or less is measured, a short circuit
check error occurs.
When [SHORT CHECK TIME] is set to [AUTO] and the equipment to be
measured is short-circuited, it takes 0.5 s for a short circuit check and “SHORT
CHECK 0.500 s” is displayed.
Setting to MANUAL
1
Open the Settings screen.
Select [SET].
2
Set [SHORT CHECK] to [ON].
(When [ON] is selected, the [SHORT CHECK TIME] setting is
displayed below.)
Selection
[ON]
[OFF]
Return to the
Measurement
screen.
3
Set [SHORT CHECK TIME] to [MANUAL].
Selection
[MANUAL]
Return to the
Measurement
screen.
10
9
8
7
6
5
4
3
2
1