EasyManua.ls Logo

Hitachi TM3000 - File Menu

Hitachi TM3000
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 - 43
NOTE : The scaling cannot be set in the hide mode. The observation mode (see Section
4.3.6 5. ) is display when the charge-up reduction mode is selected in an
observation-enabled status (when the specimen chamber is in a vacuum state).
The Image mode (see Section 4.3.5 2. ) is displayed in non-standard modes.
4.3.3 File Menu
Figure 4.3.3-1 File Menu
1. Observation
Figure 4.3.3-2 File Menu (Observation Menu)
1a. Start (HV-ON) To start the observation process, turn on the electron gun (HV-ON)
and apply a current to the filament. During this process, which takes approximately 10
seconds, a process-in-progress window appears. The same operation can also be performed
by pressing the Start button.
NOTE
: This operation cannot be conducted when the Start button is in the [Stop] (HV-ON)
status.
1b. Stop (HV-OFF) To terminate observation, turn off the electron gun (HV-OFF), and
shut off the current to the filament. The same operation can also be performed by pressing
the Stop button.
NOTE
: This operation cannot be conducted when the Start button is in the [Start]
(HV-OFF_) status.

Table of Contents

Related product manuals