Code Error Message Countermeasure
1265 The image is being observed.
Do you shut down the
application?
When shutdown of the application was tried
during electron beam irradiation (HV-ON status),
this message will open to confirm your intention
of shutdown. If you click the [OK] button, the
electron beam will be turned off and the
application will be shutdown. If you click the
[Cancel] button, the shutdown will be canceled.
1301 Time has run out.
“Auto B/C Adjustment” cannot
be executed.
This message appears if a time-out (50 s) has
occurred during auto B/C adjustment. Auto B/C
adjustment stops, and brightness and contrast
are reset to the levels before this adjustment. If
auto B/C adjustment remains unsuccessful,
brightness and contrast should be adjusted
manually.
1311 Time has run out.
“Auto Focus” cannot be
executed.
This message appears if a time-out (50 s) has
occurred detected during auto focus. Auto focus
stops, and focus condition is reset to the level
before auto focus. If auto focus remains
unsuccessful, manual focus adjustment is
required.
1390 The focus position is not within a
proper range.
Adjust the specimen stub height
to proper height.
This message appears when the following
condition lasts for 2 minutes; electron beam is
projected (HV-ON status) and focus position is
not proper because the specimen holder position
is too low. Click the [OK] button to dismiss the
message. Stop image observation once and
readjust the height of specimen holder properly.
For specimen preparation and height adjustment,
refer to 4.2.2 in the instruction manual.
1395 The focus position is not within
the proper range.
Adjustment of the specimen
stub height is recommended. Do
you continue the processing?
This confirmation message appears if the focus
position is not proper for this operation (causing
a large magnification error) while trying to
perform quick save, standard save or digital
zoom.
Clicking the [OK] the button continues the current
process; clicking the [Cancel] button cancels the
process.
It is recommended to stop image observation
once and readjust the height of specimen holder
properly. For specimen preparation and height
adjustment, refer to 4.2.2 in the instruction
manual.