EasyManua.ls Logo

Hitachi TM3000 - Page 260

Hitachi TM3000
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 - 43
Code Error Message Countermeasure
1575 The power supply of +18V for
C2 was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F14 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1576 The power supply of +18V for
C1 was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F13 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1577 The power supply of +18V for
OBJ was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F12 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1578 The power supply of -15V for
STG was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F11 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1579 The power supply of +15V for
STG was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F10 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1580 The power supply of -15V for
DEF was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F9 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.

Table of Contents

Related product manuals