79
(c) Loosen the set screw, remove the two screws of the nozzle orifice, and remove the nozzle orifice.
(d) Put the removed nozzle orifice into the beaker filled with makeup and clean it.
With the nozzle orifice removed, clean the nozzle with makeup as shown in the figure below.
Install the nozzle orifice according to Steps (a) to (g).
(a) If the O-ring comes off together with the nozzle orifice, hold the ring
with the tweezers and insert it into the nozzle.
(b) Pour a few drops of makeup into the O-ring from the cleaning bottle to
make it easier to insert the nozzle orifice.
Screw
Tweezers
O-ring
(If the O-ring comes
off together with the
nozzle orifice, store
the ring in the
beaker filled with
makeup to prevent
loss of it.)
Nozzle orifice (Do not touch the ejection port.)
2
Cleaning bottle
3
Tweezers
O-ring
O-ring
Cleaning bottle