7. On the Biopsy Control Module, select the correct Lateral Arm Mounting Side button
for the side the Lateral Arm is mounted on.
Figure 26: Lateral Arm Mounting Side Selection Screen
8. On the Biopsy Control Module, select OK to confirm that the Lateral Arm Stand is
installed.
Removal of the Lateral Arm
1. On the standard device mount, turn the device mount knob to loosen the device
mount screw.
2. Pull the Lateral Arm away from the Biopsy Guidance Module.
3. Remove the Lateral Arm Paddle.
4. Remove the Lateral Arm Stand.
Needle Guide for Lateral Arm
Warning:
Always use sterile techniques when you use Needle Guides during the
patient procedures.
Warning:
It is important to install the device correctly. Be sure to insert the needle
through the top and bottom Needle Guides.
Note