Affirm Breast Biopsy Guidance System Service Manual
Chapter 5: Care and Cleaning
Page 62 MAN-05754 Revision 002
5.2 For General Cleaning
Use a lint-free cloth or pad and apply a diluted dishwashing liquid.
Caution:
Use the least possible amount of cleaning fluids. The fluids must not flow or run.
If more than soap and water is required, Hologic recommends any one of the following:
10% chlorine bleach and water with one part commercially available chlorine bleach
(normally 5.25% chlorine and 94.75% water) and nine parts water
Commercially available isopropyl alcohol solution (70% isopropyl alcohol by
volume, not diluted)
3% maximum concentration of hydrogen peroxide solution
After you apply any of the above solutions, use a pad and apply a diluted dishwashing
liquid to clean any parts which touch the patient.
Warning:
If a paddle touches possible infectious materials, contact your Infection
Control Representative to remove contamination from the paddle.
Caution:
To prevent damage to the electronic components, do not use disinfectant sprays on
the system.
5.2.1 How to Clean the Biopsy Control Module Screen
There are many commercially available products to clean LCD screens. Make sure the
product that you select is free of strong chemicals, abrasives, bleach, and detergents that
contain fluorides, ammonia, and alcohol. Follow the directions of the manufacturer of the
product.