Affirm Breast Biopsy Guidance System Service Manual
Chapter 3: Installation, Verification, and Removal
MAN-05754 Revision 002 Page 13
3: Installation, Verification, and Removal
3.1 Add Licensed Features
Install the license(s) for the Affirm breast biopsy guidance system and/or optional Lateral
Arm upright biopsy accessory. When installing a new licensed feature on a system,
follow these steps in System (Service) Tools.
1. Obtain the Sales Order number for that feature.
2. Launch System (Service) Tools:
If there is a shortcut to System Tools on the Windows desktop, launch System
Tools and continue at step 4.
If there is no shortcut on the Windows desktop, continue at step 3.
3. Access System Tools from the Capture application:
a. Launch the Capture application and log in as a Service user.
b. At the Select Patient screen, select the Admin button.
c. At the Admin screen, select System Tools.
Note
If you see an Admin rights warning after logging in to the Capture application, you
need to restart Windows and log in again. Otherwise, you won't have permissions to
run System Tools.