Affirm Breast Biopsy Guidance System Service Manual
Chapter 4: Calibration Procedures
MAN-05754 Revision 002 Page 45
4: Calibration Procedures
This chapter describes the tools and procedures required to perform the Biopsy
subsystem calibrations.
The Guidance Hardware calibrations do not involve imaging and can be implemented by
using CalTool. The Geometry and STX calibrations require image capture.
4.1 Required Tools and Equipment
5 cm x 5 cm Standard Biopsy Paddle
ASY-03667
6 cm x 7 cm Standard Biopsy Paddle
ASY-03668
5 cm x 5 cm Axillary Biopsy Paddle
ASY-03669
QAS Phantom
ASY-06255
QAS Needle Assembly
ASY-03949
Calibration Phantom (Geometry Cal)
ASY-02042
X-ray and Light Field Fixture
TLS-01063
Dowel Pin
MME-01182