Affirm Breast Biopsy Guidance System Service Manual
Chapter 3: Installation, Verification, and Removal
MAN-05754 Revision 002 Page 37
4. Slide the disposable needle guide (item 1) over the pin on the end of the needle guide
holder (item 2).
Figure 29: Installing Disposable
Needle Guide on Needle Guide Holder
(Lateral Arm)
Figure Legend
1. Disposable Needle Guide
2. Needle Guide Holder
Removal of the Disposable Needle Guide
1. Remove the biopsy device and adapter from the device mount.
2. Pull the disposable needle guide away from the pin on the needle guide holder.
3. Discard the disposable needle guide in accordance with local regulations.
Device Mount for Lateral Arm
Installation of the Device Mount
1. Flip the Carriage lever completely up to the unlocked position.
Figure 30: Carriage Lever Locked and Unlocked Positions
Note
The Carriage lever must be fully in the up position for smooth movement along the
Lateral Arm.