Faxitron CT Specimen Radiography System User Guide
Chapter 5: Calibration
Page 40 5081-9544 Revision 004
Figure 33: Calibration Menu
Ensure the chamber is empty, including the foam tray, check the box to confirm the
chamber is empty, then click the Start button.
Figure 34: Start Image Calibration
A Calibration Progress Window will open, and the system will begin Image Calibration.
Figure 35: Calibration Progress Window
X-Rays are generated as part of the warmup and calibration procedures. If the user needs
to stop the calibration, they may click the Cancel button in the Calibration Progress