Faxitron CT Specimen Radiography System User Guide
Chapter 7: Quality Control
5081-9544 Revision 004 Page 81
7: Quality Control
7.1 Required Quality Control Procedures
The following procedures are necessary for correct system operation.
Table 1: Required Procedures
Procedures Frequency
Instructions
Image Calibration Daily Section 5.1
Geometric Calibration Quarterly Section 5.2
Uniformity Calibration Quarterly Section 5.3
7.2 Service and Troubleshooting
7.2.1 Service
There are no serviceable parts in the Faxitron CT. Refer to the copyright page at the front
of this manual.
7.2.2 Troubleshooting
The following are a few issues that you may be able to solve with some basic
troubleshooting steps.
Computer or System will not turn on.
No power to the computer or X-Ray system.
Verify power to system is ON and power cables are plugged in and the rear power
switch is set to (I) ON.
System turned off when acquiring 3D image.
The must be on mains power to acquire an image. If not, the power could overload
the UPS and shut down.
Check to make sure the system is connected to mains and the power inlet is switched
on.
There appears to be another object in the image.
Verify that during “Calibration” there were no objects in the X-ray chamber,
including the disposable tray.
Remove objects and repeat Calibration.
Poor image quality.
If artifacts are seen in the image complete a full Calibration.