EasyManua.ls Logo

Huawei NetCol5000-A035 - Page 231

Huawei NetCol5000-A035
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 6-21 Replacing the humidier cylinder
Step 4 Loosen the belt around the humidier cylinder and remove the cylinder, as shown
by (2) and (3) in Figure 6-21.
Step 5 Open the rubber pad on the cable port, and clear the glue on the port using a
at-head screwdriver, as shown by (4) in Figure 6-21.
Step 6 Remove the screws from the electrode cable connection terminals above the
humidier cylinder using the 10# socket wrench, remove the cable connection
terminals, and remove the cable from the high water level sensor, as shown by (5)
in Figure 6-21.
Step 7 Install a new humidier cylinder by performing Step 3 to Step 6 in the reverse
order, and start the humidier.
NO TE
When replacing the humidier cylinder, you do not need to apply glue to the cable
connection port again.
Step 8 Power on the
humidier, choose Maint > Diagnostic mode on the main menu,
and enter 100%. The water is injected to the
humidier. Check whether the
humidier works properly. If yes, exit the diagnostic mode.
NetCol5000-A(025, 035) In-row Air Cooled Smart
Cooling Product
User Manual (600 mm Width) 6 System Operation and Maintenance
Issue 17 (2020-10-22) Copyright © Huawei Technologies Co., Ltd. 221

Table of Contents

Other manuals for Huawei NetCol5000-A035

Related product manuals