9 Adjusting Settings
Correction Factor
e Bolus Calculator uses your Correction Factor to calculate a correction bolus
when your blood glucose value is above your Correct Above setting (see “e
Bolus Calculator” on page 143). To edit this value:
1. Write a list of all of the desired Correction Factor segments to guide you
through re-entering the values for each segment.
Tip: You can write this list on the pages at the end of this User Guide.
2. Navigate to: Settings icon ( ) > Bolus > Correction Factor.
3. Tap NEXT.
4. To edit a segment, tap the row containing the segment you want to edit.
a. Tap the End Time eld to enter a new end time.
b. Tap the 1 Unit of Insulin decreases BG by eld and enter a new
Correction Factor.
c. Tap NEXT.
5. Repeat the previous step as needed for the remaining segments.
6. Aer conrming that all segments are correct, tap SAVE.
Note: You can add and delete segments by editing the existing segments.
Reverse Correction
e Reverse Correction setting determines how the Bolus Calculator handles
meal boluses when your blood glucose value is below your Target BG (see
“Reverse Correction” on page 145 for more details).
To turn Reverse Correction on or o:
1. Navigate to: Settings icon ( ) > Bolus.
2. Tap the toggle on the Reverse Correction line to turn it on or o.
Duration of Insulin Action
e Bolus Calculator uses your Duration of Insulin Action setting to calculate the
amount of insulin on board (IOB) from a previous bolus. To edit this value:
1. Navigate to: Settings icon ( ) > Bolus > Duration of Insulin Action.
2. Tap the Duration of Insulin Action eld and enter the new value.
3. Tap SAVE.