12 Understanding PDM and Pod Function
When the Bolus Calculator does not work
e Bolus Calculator does not work when it is disabled or when it is set to o. You
control whether the Bolus Calculator is turned on or o, but the PDM controls
when it is disabled.
If the Bolus Calculator is set to o, the PDM does not track IOB and does not
suggest a bolus.
If the Bolus Calculator is set to on, the PDM may disable it in a few situations.
Being disabled means that the Bolus Calculator is temporarily unable to calculate
a suggested bolus.
Conditions that disable
the Bolus Calculator:
e Bolus Calculator is
disabled until:
While the Bolus
Calculator is
disabled:
Your blood glucose
reading is below your
Minimum BG for Calcs
setting.
Ten minutes pass.
or
A new blood glucose reading
is above your Minimum BG
for Calcs setting.
IOB is displayed on
the Home screen.
Your blood glucose
reading is “HI”.
Ten minutes pass.
or
A new blood glucose reading
is lower than “HI”.
IOB is displayed on
the Home screen.
ere is an unconrmed
bolus when you discard
a Pod.
A complete Duration of
Insulin Action period passes.
IOB is not displayed
on the Home screen.
ere is an internal
clock reset.
8.5 hours pass. IOB is not displayed
on the Home screen.
Factors used in the Bolus Calculator’s calculations
e Bolus Calculator accounts for the following when it calculates a bolus:
• Your current blood glucose level, Target BG, Correct Above threshold and
Correction Factor
• e carbs you are about to eat and your IC Ratio
• Your Duration of Insulin Action and insulin on board (IOB)
• Your Minimum BG for Calcs
• Reverse Correction, if it is enabled