Frequency: 2,400-2,480 GHz, digitally modulated, with an eective radiated
power of 1.14 mW
e Omnipod DASH™ System complies with the immunity requirements of the
general standard for electromagnetic compatibility, IEC 60601-1-2.
e Omnipod DASH™ System has demonstrated immunity to both radio-
frequency identication (RFID) systems and Electronic Article Surveillance
(EAS) systems. Testing has been performed in accordance with AIM 7351731 to
demonstrate this immunity.
Caution: Portable RF communications equipment (including peripherals, such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Omnipod DASH™ System. Otherwise, degradation
of the performance of this equipment could result.
Electromagnetic Emissions
is equipment is intended for use in the electromagnetic environment specied below. e
user of this equipment should assure that it is used in such an environment.
Emissions Compliance
according to
Electromagnetic environment
RF Emissions
(CISPR 11)
Group 2 e Pod and the PDM emit low-level
electromagnetic energy (RF) in order to
communicate. Although unlikely, nearby
electronic equipment may be aected.
CISPR B
Emissions
Classication
Class B e System is suitable for use in all establishments
including domestic establishments.
Electromagnetic Immunity
e System is intended for use in the electromagnetic environment specied
below. You should observe these requirements when using the System.
Immunity
against
IEC 60601-1-2
test level
Compliance
level (of this
device) Electromagnetic environment
ElectroStatic
Discharge, ESD
(IEC 61000-4-2)
contact
discharge:
± 8 kV
air discharge:
± 15 kV
± 8 kV
± 15 kV
If oors are covered with synthetic material,
try to avoid electrostatic discharges.
Power
frequency
magnetic elds
50/60 Hz (IEC
61000-4-8)
30 A/m 400 A/m Suitable for most environments. Magnetic
eld strengths in excess of 400 A/m would
be unlikely except in close proximity to
industrial magnetic devices.
Effective Date: 27SEP2019, ECO-000240