EasyManua.ls Logo

Insulet Omnipod Dash - Page 201

Insulet Omnipod Dash
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
184
Appendix
Frequency: 2,400-2,480 GHz, digitally modulated, with an eective radiated
power of 1.14 mW
e Omnipod DASH™ System complies with the immunity requirements of the
general standard for electromagnetic compatibility, IEC 60601-1-2.
e Omnipod DASH™ System has demonstrated immunity to both radio-
frequency identication (RFID) systems and Electronic Article Surveillance
(EAS) systems. Testing has been performed in accordance with AIM 7351731 to
demonstrate this immunity.
Caution: Portable RF communications equipment (including peripherals, such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Omnipod DASH™ System. Otherwise, degradation
of the performance of this equipment could result.
Electromagnetic Emissions
is equipment is intended for use in the electromagnetic environment specied below. e
user of this equipment should assure that it is used in such an environment.
Emissions Compliance
according to
Electromagnetic environment
RF Emissions
(CISPR 11)
Group 2 e Pod and the PDM emit low-level
electromagnetic energy (RF) in order to
communicate. Although unlikely, nearby
electronic equipment may be aected.
CISPR B
Emissions
Classication
Class B e System is suitable for use in all establishments
including domestic establishments.
Electromagnetic Immunity
e System is intended for use in the electromagnetic environment specied
below. You should observe these requirements when using the System.
Immunity
against
IEC 60601-1-2
test level
Compliance
level (of this
device) Electromagnetic environment
ElectroStatic
Discharge, ESD
(IEC 61000-4-2)
contact
discharge:
± 8 kV
air discharge:
± 15 kV
± 8 kV
± 15 kV
If oors are covered with synthetic material,
try to avoid electrostatic discharges.
Power
frequency
magnetic elds
50/60 Hz (IEC
61000-4-8)
30 A/m 400 A/m Suitable for most environments. Magnetic
eld strengths in excess of 400 A/m would
be unlikely except in close proximity to
industrial magnetic devices.
Effective Date: 27SEP2019, ECO-000240