SERVICE MANUAL
Invacare
®
- Bora/Spectra XTR
58
CAUTION!
Property damage can result if the counter is wrongly connected. If the plus and minus
wires are connected the wrong way, it will damage the electronic components of the
operating hours counter.
Make sure the cable is connected correctly.
The operating hours counter is installed in
reverse order.
Connect the plug on the blue wire (-) to the
pin on the operating hours counter which is
marked with a “minus” symbol.
Connect the plug on the brown wire (+) to the
pin on the operating hours counter which is
marked with a “plus” symbol.
The brown wire is additionally marked with a
yellow sleeve with a “plus” symbol on it.
Close the casing and reattach it to the
wheelchair with the Velcro fastening.
Reinstall all remaining components in reverse
order.

Table of Contents