EasyManua.ls Logo

Keithley 2016 - Page 153

Keithley 2016
269 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 5-6
ROUTe command summary
Command Description
Default
parameter
SCPI
:ROUTe
:SCAN
:EXTernal <list>
:EXTernal?
:LSELect <name>
:LSELect?
Commands to control scanner card:
Path to scan channels.
Specify external scan list (2 to 800 channels).
Query external scan list.
Select scan operation (EXTernal or NONE).
Query scan operation.
1-10
NONE
Table 5-7
SENSe command summary
Command Description
Default
parameter
SCPI
[:SENSe[1]]
:FUNCtion <name>
:FUNCtion?
:DATA?
:DATA
:FRESh?
:LATest?
:HOLD
:WINDow <NRf>
:WINDow?
:COUNt <NRf>
:COUNt?
:STATe <NRf>
:STATe?
Select measurement function: ‘VOLTage:AC’, VOLTage
:DC’, RESistance’, ‘FRESistance’, ‘CURRent:AC’,
‘CURRent: DC’, ‘FREQuency’, ‘TEMPerature’, ‘PERiod’,
‘DIODe’, ‘CONTinuity’, ‘DISTortion’
Query function.
Return the last instrument reading.
Return latest reading after trigger (similar to FETch?)
Returns last reading the instrument had.
Path to control Hold feature:
Set Hold window (%); 0.01 to 20.
Query Hold window.
Set Hold count; 2 to 100.
Query Hold count.
Enable or disable Hold.
Query state of Hold.
‘VOLT[:DC]’
1
5
OFF
SCPI Command Reference 5-11
Section 5A.fm Page 11 Thursday, December 11, 2003 8:22 AM

Table of Contents