:CURRent:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
:DETector
:BANDwidth <NRf>
:BANDwidth?
Path to congure AC current.
Set integration rate (line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to congure measurement range:
Select range (0 to 3.1).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (-3.1 to 3.1).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 7).
Query resolution.
Path to congure and control the lter.
Select lter type: (MOVing or REPeat).
Query lter type.
Specify lter count (1 to 100).
Query lter count.
Enable or disable lter.
Query state of digital lter.
Path to congure bandwidth:
Specify bandwidth (3 to 300e3).
Query bandwidth.
1
3
ON
0
OFF
6
(Note)
10
OFF
30
:CURRent:DC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
Path to congure DC current.
Set integration rate (line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to congure measurement range:
Select range (0 to 3.1).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (-3.1 to 3.1).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 7).
Query resolution.
Path to congure and control the lter.
Select lter type: (MOVing or REPeat).
Query lter type.
Specify lter count (1 to 100).
Query lter count.
Enable or disable lter.
Query state of digital lter.
1
3
ON
0
OFF
7
(Note)
10
OFF
Table 5-7 (cont.)
SENSe command summary
Command Description
Default
parameter
SCPI
5-12 SCPI Command Reference