Section 10: Triggering Series 2600A System SourceMeter® Instruments Reference Manual
Output characteristics:
• In addition to trigger events from other trigger objects, the digio.trigger[N].assert
and tsplink.trigger[N].assert commands generate a low pulse for the
programmed pulse duration if the line is latched low, a falling edge does not occur.
• A normal falling edge pulse generates when the internal drive is not latched low and the
digio.trigger[N].assert and tsplink.trigger[N].assert commands are
issued.
• To mirror the SynchronousA trigger mode, set the pulse width to 1μs or any small nonzero
value.
• Action overruns are disabled.