18-4 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 18: TSP-NetTM Series 2600A System SourceMeter® Instruments Reference Manual
These advantages make using TSP Link to control another TSP-enabled device the best choice
for most applications. However, if the distance between the Series 2600A and the TSP-enabled
device is longer than 15 feet, use TSP-Net.
To establish a remote TSP-Net connection with a TSP-enabled device, use tspnet.connect
without specifying a port number. The Series 2600A enables TSP prompt and error handling for
the remote device, which allows you to successfully use the tspnet.tsp set of commands to load
and run scripts and retrieve reading buffers.
Abort any operation on the remote TSP-enabled device using abort().