19-84 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
lan.trigger[N].stimulus
Replace N with values 1-8.
Attribute Selects which events will trigger a LAN trigger packet.
Default 0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage stimulus = lan.trigger[lanevent].stimulus -- Reads trigger identifier.
lan.trigger[lanevent].stimulus = stimulus -- Writes trigger identifier.
lanevent The LAN event number.
stimulus Identifier for the triggering event.
Remarks Setting this attribute to zero will disable automatic trigger generation.
If any events are detected prior to calling lan.trigger[n].connect, the event will be
ignored and the action overrun will be set.
lan.trigger[N].wait
Replace N with values 1-8.
Function Wait for LAN trigger event to be received.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage triggered = lan.trigger[lanevent].wait(timeout)
lanevent The LAN event number.
timeout Maximum amount of time (in seconds) to wait for
the trigger event.
triggered Trigger detected indicator.
Remarks If one or more trigger events have been detected since the last time lan.trigger[n].wait
or lan.trigger[n].clear was called, this function will return immediately.
After waiting for a LAN trigger event, the event detector will be automatically reset regardless of
the number of events detected.

Table of Contents