19-242 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
Parameters
smu, startv, stopv, stime, points
smu: SourceMeter instrument channel (A or B). Defaults to SMU A if
all parameters are omitted when function is called.
starti: Sweep start voltage in volts.
stopi: Sweep stop voltage in volts.
stime: Settling time in seconds. Occurs after stepping the source
and before performing a measurement.
points: Number of sweep points (must be 2).
Data
Current measurements, voltage source values and timestamps are stored in
smuX.nvbuffer1.
Example
SweepVLinMeasureI(smua, -1, 1, 1E-3, 1000)
This function performs a 1000-point linear voltage sweep starting at -1V and
stopping at +1V. Current is measured at every step (point) in the sweep after a
1ms source settling period.
Table 19-4 (cont.)
KISweep TSP test script: SweepVLinMeasureI
TSP project name: KISweep
TSP test script name: KISweep