EasyManuals Logo

Keithley 2600a series User Manual

Keithley 2600a series
696 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #564 background imageLoading...
Page #564 background image
19-242 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
Parameters
smu, startv, stopv, stime, points
smu: SourceMeter instrument channel (A or B). Defaults to SMU A if
all parameters are omitted when function is called.
starti: Sweep start voltage in volts.
stopi: Sweep stop voltage in volts.
stime: Settling time in seconds. Occurs after stepping the source
and before performing a measurement.
points: Number of sweep points (must be ≥2).
Data
Current measurements, voltage source values and timestamps are stored in
smuX.nvbuffer1.
Example
SweepVLinMeasureI(smua, -1, 1, 1E-3, 1000)
This function performs a 1000-point linear voltage sweep starting at -1V and
stopping at +1V. Current is measured at every step (point) in the sweep after a
1ms source settling period.
Table 19-4 (cont.)
KISweep TSP test script: SweepVLinMeasureI
TSP project name: KISweep
TSP test script name: KISweep

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals