19-244 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
Example SweepILogMeasureV(smua, 0.01, 0.1, 0.001, 5)
This function performs a 5-point logarithmic current sweep starting at 10mA and
stopping at 100mA. Voltage is measured at every step (point) in the sweep. The
source will be allowed to settle on each step for 1ms before a measurement is
performed.
The following log values and corresponding source levels for the 5-point log
sweep are listed as follows:
MeasurePoint LogStepSize LogStep SourceStepLevel
1 0.25 0.0 0.01A
2 0.25 0.25 0.017783A
3 0.25 0.5 0.031623A
4 0.25 0.75 0.056234A
5 0.25 1.0 0.1A
Table 19-5 (cont.)
KISweep TSP test script: SweepILogMeasureV
TSP project name: KISweep
TSP test script name: KISweep
Function:
SweepILogMeasureV(smu, starti, stopi, stime, points)