u. Send the source calibration command using the -range, -zero and -FS multimeter read-
ings, and -zero and -FS source values for the parameters:
smua.source.calibratei(-range,src_Z,DMM_Z_rdg,
src_FS,DMM_FS_rdg)
Where: -range -- negative of the present calibration range
src_Z -- zero 2600A source output value
DMM_Z_rdg -- zero DMM measurement
src_FS -- FS 2600A source output value
DMM_FS_rdg -- FS DMM measurement
Typical values for the 1A range:
smua.source.calibratei(-1,-1e-10,-1e-5,-0.9,-0.892)
v. If this step is not on the CALA sense mode, send the measure calibration command
using the multimeter and Series 2600A readings, and range setting for the parameters:
smua.measure.calibratei(-range,Z_rdg,DMM_Z_rdg,
FS_rdg,DMM_FS_rdg)
Where: -range -- negative of the present calibration range
Z_rdg -- zero 2600A measurement
DMM_Z_rdg -- zero DMM measurement
FS_rdg -- FS 2600A measurement
DMM_FS_rdg -- FS DMM measurement
Typical 1A range values:
smua.measure.calibratei(-1,-1e-4,-1e-5,-0.91,-0.892)
4. Be sure to complete steps a through v for the 100nA to 1A ranges before continuing with 3A
and 10A range calibration (Model 2601A/2602A) or 1.5A and 10A range calibration (Model
2611A/2612A/2635A/2636A).
5. Change connections as shown in Figure 20-3 (use 4-wire connections to the 0.5Ω resistor
as shown).
6. Select the DMM DC volts function.
7. Repeat steps a through v for the 3A and 10A ranges (Model 2601A/2602A) or 1.5A and 10A
ranges (Model 2611A/2612A/2635A/2636A). Compute the current reading from the DMM
voltage reading and characterized 0.5Ω resistance value: I = V/R.
8. Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO