2600AS-901-01 Rev. B / September 2008 Return to Section Topics C-5
Series 2600A System SourceMeter® Instruments Reference Manual Appendix C: Status Model
Figure C-3
Status model (operation event registers)
Bit
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status
Register (operation/OSR)
+
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status
Calibration Summary
+
SMU A (SMUA)
SMU B (SMUB)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status
Instrument Summary
Register
+
To Operation Summary Bit
(OSB) in Status Byte
(status.condition/STB)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status SMU A
Summary Register
+
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status SMU B
Summary Register
+
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status User
Register
+
Similar registers
exists for Measuring
and Sweeping
status.operation
status.operation.calibrating
(status.operation.measuring)
(status.operation.sweeping)
status.operation.instrumentstatus.operation.instrument.smua
status.operation.instrument.smub
status.operation.user
Bit 0 (BIT0)
Bit 1 (BIT1)
Bit 6 (BIT6)
Bit 3 (BIT3)
4 (BIT4)
Bit 5 (BIT5)
Bit 2 (BIT2)
Bit 7 (BIT7)
Bit 8 (BIT8)
Bit 9 (BIT9)
Bit 10 (BIT10)
Bit 11 (BIT11)
Bit 12 (BIT12)
Bit 13 (BIT13)
Bit 14 (BIT14)
Program Running (PROG)
Instrument Summary (INST)
User (USER)
Remote Summary (REM)
Trigger Overrun (TRGOVR)
Waiting for Trigger
Measuring (MEAS)
Sweeping (SWE)
Calibrating (CAL)
Calibrating (CAL)
Waiting for Trigger
Measuring (MEAS)
Sweeping (SWE)
Trigger Overrun (TRGOVR)
SMU A (SMUA)
Digital I/O (digio)
SMU B (SMUB)
LAN (LAN)
TSP-Link (TSPLINK)
Digital I/O (DIGIO)
Trigger Timer (TRGTMR)
Trigger Blender (TRGBLND)
Calibrating (CAL)
Waiting for Trigger
Measuring (MEAS)
Sweeping (SWE)
Trigger Overrun (TRGOVR)

Table of Contents