9-24 Sweep Operation
‘Setup Source Memory Location 2
‘------------------------------
Call OutputCmd(intGPIB, “*RST”) ‘Restore GPIB default conditions.
Call OutputCmd(intGPIB, “:SOUR:FUNC VOLT”) ‘Current Source Function.
Call OutputCmd(intGPIB, “:SENS:FUNC ‘CURR:DC’”) ‘Current Sense Function.
Call OutputCmd(intGPIB, “:SENS:CURR:PROT .105”) ‘Set 105mA Compliance
Call OutputCmd(intGPIB, “:SENS:CURR:RANGE .1”) ‘Set 100mA Current Measure Range
Call OutputCmd(intGPIB, “:SOUR:DEL 1”) ‘Set Source Delay to 1
Call OutputCmd(intGPIB, “:SOUR:VOLT 10”) ‘10V Source Voltage.
Call OutputCmd(intGPIB, “:CALC2:FEED CURR”) ‘Send Current(A) Readings to Buffer
Call OutputCmd(intGPIB, “:CALC2:LIM1:STAT ON”) ‘Set Limit 1 on
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:FAIL IN”) ‘Set Fail Mode to In Compliance
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:SOUR2 5”) ‘Set Digital Output Pattern for Compliance Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:STAT ON”) ‘Set Limit2 on
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP 3E-2”) ‘Set Upper Limit to 30mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW 2E-2”) ‘Set Lower Limit to 20mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP:SOUR2 4”) ‘Set Digital Output Pattern for Upper Limit #2
Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW:SOUR2 5”) ‘Set Digital Output Pattern for Lower Limit #2
Failure
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SOUR2 14”) ‘Set Digital Output Pattern for Limit #2 Passing
Call OutputCmd(intGPIB, “:CALC2:CLIM:MODE GRAD”) ‘Set Limit Results to Grading
Call OutputCmd(intGPIB, “:CALC2:CLIM:BCON END”) ‘Set Binning Control to End
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SML NEXT”) ‘Set SML Pass Location
Call OutputCmd(intGPIB, “:CALC2:CLIM:FAIL:SML 100”) ‘Set SML Fail Location
Call OutputCmd(intGPIB, “:SOUR:MEM:SAVE 2”) ‘Save in Source Memory Location 2.
‘Setup Source Memory Location 3
‘------------------------------
Call OutputCmd(intGPIB, “*RST”) ‘Restore GPIB default conditions.
Call OutputCmd(intGPIB, “:SOUR:FUNC VOLT”) ‘Current Source Function.
Call OutputCmd(intGPIB, “:SENS:FUNC ‘CURR:DC’”) ‘Current Sense Function.
Call OutputCmd(intGPIB, “:SENS:CURR:PROT .105”) ‘Set 105mA Compliance
Call OutputCmd(intGPIB, “:SENS:CURR:RANGE .1”) ‘Set 100mA Current Measure Range
Call OutputCmd(intGPIB, “:SOUR:DEL 1”) ‘Set Source Delay to 1
Call OutputCmd(intGPIB, “:SOUR:VOLT 10”) ‘10V Source Voltage.
Call OutputCmd(intGPIB, “:CALC2:FEED CURR”) ‘Send Current(A) Readings to Buffer
Call OutputCmd(intGPIB, “:CALC2:LIM1:STAT ON”) ‘Set Limit 1 on
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:FAIL IN”) ‘Set Fail Mode to In Compliance
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:SOUR2 10”) ‘Set Digital Output Pattern for Compliance Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:STAT ON”) ‘Set Limit2 on
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP 4E-2”) ‘Set Upper Limit to 40mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW 3E-3”) ‘Set Lower Limit to 30mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP:SOUR2 6”) ‘Set Digital Output Pattern for Upper Limit #2
Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW:SOUR2 7”) ‘Set Digital Output Pattern for Lower Limit #2
Failure
Call OutputCmd(intGPIB, “:CALC2:cLIM:PASS:SOUR2 15”) ‘Set Digital Output Pattern for Limit #2 Passing
Call OutputCmd(intGPIB, “:CALC2:CLIM:MODE GRAD”) ‘Set Limit Results to Grading
Call OutputCmd(intGPIB, “:CALC2:CLIM:BCON END”) ‘Set Binning Control to End
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SML NEXT”) ‘Set SML Pass Location
Call OutputCmd(intGPIB, “:CALC2:CLIM:FAIL:SML 100”) ‘Set SML Fail Location
Call OutputCmd(intGPIB, “:SOUR:MEM:SAVE 3”) ‘Save in source memory location 3.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω