19-22 Calibration
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be acces-
sible in the mainframe calibration menu. To view calibration dates and calibration count:
1. From normal display, press MENU, select CAL, and then press ENTER. The unit will
display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
2. Select VIEW-DATES, and then press ENTER. The Model 6430 will display the next
and last calibration dates and the calibration count as in the following example:
NEXT CAL: 12/15/1999
Last calibration: 12/15/1999 Count: 0001