EasyManuals Logo

Keithley 6430 User Manual

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #408 background imageLoading...
Page #408 background image
17-98 SCPI Command Reference
Reset timestamp
:RESet
:SYSTem:TIME:RESet Reset timestamp
Description This action command is used to reset the absolute timestamp to 0 seconds.
The timestamp also resets to 0 seconds every time the SourceMeter is turned
on.
Auto reset timestamp
:RESet:AUTO <b>
:SYSTem:TIME:RESet:AUTO <b> Reset timestamp when exiting idle
Parameters <b> = 1 or ON Enable auto timestamp reset
0 or OFF Disable auto timestamp reset
Query :AUTO? Query enabled/disabled auto timestamp reset
state
Description :RES:AUTO enables or disables auto timestamp reset. When enabled, the
timestamp will be automatically reset when exiting the idle layer of the trig-
ger model. This command is intended for use with READ?/INIT when tak-
ing more than one reading.
Auto range change mode
:RCMode <name>
:SYSTem:RCMode <name> Control auto range change mode
Parameters <name> = SINGle Single mode
MULTiple Multiple mode
Query :RCMode? Query auto range change mode
Description This command controls the auto range change mode. In the SINGle mode,
the SourceMeter will auto range only after rst taking a reading. In the
MULTiple mode, the SourceMeter will auto range up on compliance in the
Delay phase of the Source-Delay-Measure cycle, thereby minimizing the
chance of a SourceMeter being in compliance in a multiple-SourceMeter
system. A SourceMeter can downrange only once a reading has been taken.
In the MULTiple mode, you can control the soak time using the
:SOUR:SOAK command. See SOURce subsystem. Note that you can use
the LLIMIT and ULIMIT commands to control auto range limits. See
SENSe1 subsystem.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω

Related product manuals