EasyManua.ls Logo

Keithley 6430 - Table 6-6 Auto Filter Settings Where NPLC = 0.11 to; Table 6-7 Auto Filter Settings Where NPLC = 1.01 to

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-14 Range, Digits, Speed, and Filters
Table 6-6
Auto filter settings where NPLC = 0.11 to 1.00
Current range Repeat count Median rank Moving count
001pA
010pA
100pA
001nA
010nA
100nA
001µA
010µA
100µA
001mA
010mA
100mA
10
10
10
10
1
1
1
1
1
1
1
1
5
3
1
0
0
0
0
0
0
0
0
0
34
15
5
5
1
1
1
1
1
1
1
1
Table 6-7
Auto filter settings where NPLC = 1.01 to 10
Current range Repeat count Median rank Moving count
001pA
010pA
100pA
001nA
010nA
100nA
001µA
010µA
100µA
001mA
010mA
100mA
1
1
1
1
1
1
1
1
1
1
1
1
5
3
1
0
0
0
0
0
0
0
0
0
34
15
5
5
1
1
1
1
1
1
1
1

Table of Contents