18-6 Performance Verification
Resistance limits calculation
When verifying the resistance measurement accuracy, it will most likely be necessary to
recalculate resistance limits based on the actual calibrator resistance values. You can calculate
resistance reading limits in the same manner described above, but be sure to use the actual
calibrator resistance values and the Model 6430 normal accuracy specications for your
calculations.
As an example, assume you are testing the 20k range, and that the actual value of the nom-
inal 19k calibrator resistor is 19.01k. Using the Model 6430 20k range one-year normal
accuracy specications of ±(0.063% of reading + 3), the recalculated reading limits are:
Reading limits = 19.01k ± [(19.01k × 0.063%) + 3]
Reading limits = 19.01k ±15
Reading limits = 18.9950k to 19.0250k
Limits calculation with test equipment uncertainty
Reading limits given in this section for the 1pA-100nA and 2G-200G ranges include the
characterization accuracy of the Model 5156 Electrometer Calibration Standard. Reading limits
for these ranges are calculated as indicated above except that they also take into account Model
5156 uncertainty. For example, using the 1G Model 5156 resistor to test the Model 6430
2G range, we have:
Model 6430 normal accuracy specications: ±(0.085% + 100k)
Model 5156 1G characterization accuracy: 300ppm (0.03%)
Assuming an actual Model 5156 characterized value of 1.02G, we have:
Reading limits = 1.02G ± [(1.02G × (0.085% + 0.03%)) + 100k]
Reading limits = 1.02G ±1.273M
Reading limits = 1.01873G to 1.02127G
Performing the verication test procedures
Restoring factory defaults
Before performing the verication procedures, restore the instrument to its factory front
panel (bench) defaults as follows:
1. Press the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL
2. Select SAVESETUP, and then press ENTER. The unit then displays:
SAVESETUP MENU
GLOBAL SOURCE-MEMORY

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals