EasyManuals Logo

Keithley 6430 User Manual

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #293 background imageLoading...
Page #293 background image
Status Structure 14-15
LL3
(B3)
HL2
(B2)
LL2
(B1)
L1
(B0)
HL3
(B4)
LP
(B5)
RAV
(B6)
ROF
(B7)
BAV
(B8)
BFL
(B9)
(B10)
OT
(B12)
Measurement
Condition Register
BFL
(B9)
(B10)
Measurement Event
Register
(B10)
Measurement Event
Enable Register
&
&
&
&
&
&
OR
To Measurement
Summary Bit
(OSB) of Status
Byte Register
(Figure 14-3)
Comp = In Compliance
OVP = Over Voltage Protection
OT = Over Temperature
INT = Interlock Asserted
BFL = Buffer Full
BAV = Buffer Available
ROF = Reading Overflow
RAV = Reading Available
LP = Limits Pass
INT
(B11)
&
&
&
&
HL3 = High Limit 3
LL3 = Low Limit 3
HL2 = High Limit 2
LL2 = Low Limit 2
L1 = Limit 1
& = Logical AND
OR = Logical OR
OVP
(B13)
Comp
(B14)
(B15)
INT
(B11)
OT
(B12)
OVP
(B13)
Comp
(B14)
(B15)
INT
(B11)
OT
(B12)
OVP
(B13)
Comp
(B14)
(B15)
&
&
&
&
stat:meas:cond?
stat:meas?
stat:meas:enab <NRF>
stat:meas:enab?
L1
(B0)
LL2
(B1)
HL2
(B2)
LL3
(B3)
HL3
(B4)
LP
(B5)
RAV
(B6)
ROF
(B7)
BAV
(B8)
BFL
(B9)
BAV
(B8)
ROF
(B7)
RAV
(B6)
LP
(B5)
HL3
(B4)
LL3
(B3)
HL2
(B2)
LL2
(B1)
L1
(B0)
Figure 14-6
Measurement event status

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω