SCPI Command Reference 17-65
Select function mode
[:MODE] <name>
:SOURce[1]:FUNCtion[:MODE] <name> Select source mode
Parameters <name> = VOLTage Select voltage mode
CURRent Select current mode
MEMory Select memory mode
Query [:MODE]? Query selected source
Description This command is used to select the source mode. With VOLTage selected,
the V-Source will be used, and with CURRent selected, the I-Source will be
used.
With MEMory selected, a memory sweep can be performed. Operating set-
ups (up to 100) saved in memory can be sequentially recalled. This allows
multiple source/measure functions to be used in a sweep.
Select sourcing mode
:MODE <name>
:SOURce[1]:CURRent:MODE <name> Select DC sourcing mode for I-Source
:SOURce[1]:VOLTage:MODE <name> Select DC sourcing mode for V-Source
Parameters <name> = FIXed Select fixed sourcing mode
LIST Select list sourcing mode
SWEep Select sweep sourcing mode
Query :MODE? Query DC sourcing mode
Description This command is used to select the DC sourcing mode for the specified
source. The three modes are explained as follows:
FIXed — In this DC sourcing mode, the specified source will output a fixed
level. Use the :RANGe and :AMPLitude commands to specify the fixed
source level. See Select range and Set amplitude for fixed source.
LIST — In this mode, the source will output levels that are specified in a
list. See Configure list for commands to define and control the execution of
the list.
SWEep — In this mode, the source will perform a voltage, current or mem-
ory sweep. See Configure voltage and current sweeps and Configure mem-
ory sweep for commands to define the sweep.
NOTE The sourcing mode will default to FIXed whenever the SourceMeter goes to the local
state.