17-60 SCPI Command Reference
Set measurement speed
:NPLCycles <n>
[:SENSe[1]]:CURRent[:DC]:NPLCycles <n> Set speed (PLC)
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n> Set speed (PLC)
[:SENSe[1]]:RESistance:NPLCycles <n> Set speed (PLC)
Parameters <n> = 0.01 to 10 Power-line cycles per integration
DEFault 10
MINimum 0.01
MAXimum 10
Query :NPLCycles? Query programmed PLC value
:NPLCycles? DEFault Query *RST default PLC
:NPLCycles? MINimum Query minimum PLC
:NPLCycles? MAXimum Query maximum PLC
Description This command is used to set the integration period (speed) for measure-
ments. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC of
1, the integration period would be 1/60 (for 60Hz line power) which is 16.67
msec.
Note that this is a global command. Thus, if you set the speed for voltage
measurements to 10 PLC, then current and resistance will also set to 10
PLC.
Congure and control lters
NOTE Detailed information on the repeat, median and moving filters are provided in Sec-
tion 6,“Filters.
:AUTO <b>
[:SENSe[1]]:AVERage:AUTO <b> Enable/disable auto lter
Parameters <b> = 0 or OFF Disable auto lter
1 or ON Enable auto lter
Query :AUTO? Query state of auto lter
Description With auto lter enabled, the instrument automatically selects lter settings
for current measurements. Heavy ltering is used for the low current ranges,
and less ltering as the current range increases. The auto lter settings are
listed in Tables 6-5, 6-6, and 6-7. For voltage measurements, auto lter sets
the repeat and moving count to one, and the median rank to zero. These set-
tings effectively disable the three lters.
When auto lter is disabled, the present count and rank settings for the three
lters are used for all measurement functions and ranges.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω

Related product manuals