EasyManua.ls Logo

Keithley 6430 - 8 Data Store; Remote Command Data Store; Table 8-1 Data Store Commands

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-6 Data Store
Remote command data store
Data store commands
Table 8-1 summarizes commands associated with data store operation. See TRACe sub-
system and CALCulate3 in Section 17 for more detailed information on these commands.
Table 8-1
Data store commands
Command Description
:TRACe:DATA?
:TRACe:CLEar
:TRACe:FREE?
:TRACe:POINts <n>
:TRACe:POINts:ACTual?
:TRACe:FEED <name>
:TRACe:FEED:CONTrol <name>
:TRACe:TSTamp:FORMat <name>
:CALCulate3:FORMat <name>
:CALCulate3:DATA?
Read contents of buffer.
Clear buffer.
Read buffer memory status.
Specify buffer size (n = buffer size.)
Query number of stored readings.
Specify reading source. Name = SENSe[1] (raw readings),
CALCulate[1] (Calc1 readings), or CALCulate2 (Calc2 readings).
Start or stop buffer. Name = NEXT (fill buffer and stop) or NEVer
(disable buffer).
Select timestamp format. Name = ABSolute (reference to first buffer
reading) or DELTa (time between buffer readings).
Select buffer statistic (name = MEAN, SDEViation, MAXimum,
MINimum, or PKPK).
Read buffer statistic data.*
*If :TRACe:FEED is set to :SENSe[1], this command will return one V, I, , and MATH result.

Table of Contents

Other manuals for Keithley 6430

Related product manuals