EasyManua.ls Logo

Keithley 6430

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggering 10-23
Remote trigger commands
Table 10-1 summarizes remote trigger commands. These commands are covered in more
detail in Section 17 except for *TRG, a common command covered in Section 15.
Table 10-1
Remote trigger command
Command Description
:INITiate Take SourceMeter out of idle state.
:ABORt Abort operation, return to idle.
:ARM:COUNt <n> Set arm count (n = count).
:ARM:SOURce <name> Specify arm control source. Name = IMMediate, TLINk, TIMer, MANual,
BUS, NSTest, PSTest, or BSTest.
:ARM:TIMer <n> Set arm layer timer interval (n = interval).
:ARM:DIRection <name> Control arm bypass. (Name = SOURce or ACCeptor).
:ARM:ILINe <NRf> Select arm layer input line. (NRf = input line #).
:ARM:OLINe <NRf> Select arm layer output line (NRf = output line #).
:ARM:OUTPut <event list> Select arm layer output events. (Event list = TENTer, TEXit, or NONE).
:TRIGger:CLEar Clear any pending input triggers immediately.
:TRIGger:COUNt <n> Set trigger count (n = count).
:TRIGger:DELay <n> Set trigger delay (n = delay).
:TRIGger:SOURce <name> Specify trigger control source. Name = IMMediate or TLINk.
:TRIGger:DIRection <name> Control trigger bypass. (Name = SOURce or ACCeptor).
:TRIGger:ILINe <NRf> Select trigger layer input line (NRf = input line).
:TRIGger:OLINe <NRf> Select trigger layer output line (NRf = output line).
:TRIGger:INPut <event list> Select trigger input layer events. (Event list = SOURce, DELay, SENSe, or
NONE).
:TRIGger:OUTPut <event list> Select trigger layer output events. (Event list = SOURce, DELay, SENSe,
or NONE).
*TRG Trigger SourceMeter (if BUS source selected).

Table of Contents

Other manuals for Keithley 6430

Related product manuals