EasyManuals Logo

Keithley 6430 User Manual

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #411 background imageLoading...
Page #411 background image
SCPI Command Reference 17-101
:CONTrol <name>
:TRACe:FEED:CONTrol <name> Start or stop buffer
Parameters <name> = NEXT Fills buffer and stops
NEVer Disables buffer storage
Query :CONTrol? Query buffer control
Description This command is used to select the buffer control. When NEXT is selected,
the asterisk (*) annunciator turns on to indicate that the buffer is enabled.
The storage process starts when SourceMeter is taken out of idle to perform
source-measure operations.
After the buffer stores the specied number of reading arrays (as set by the
:POINTs command), the asterisk annuciator turns off to indicate that stor-
age is done.
With NEVer selected, storage into the buffer is disabled.
Select timestamp format
:FORMat <name>
:TRACe:TSTamp:FORMat <name> Select timestamp format
Parameters <name> = ABSolute Reference to rst buffer reading
DELTa Time between buffer readings
Query :FORMat? Query timestamp format
Description This command is used to select the timestamp format for buffer readings.
With ABSolute selected, each timestamp is referenced to the rst reading
stored in the buffer. With DELTa selected, timestamps provide the time
between each buffer reading.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω

Related product manuals