102 Keysight E5071C ENA Network Analyzers
Troubleshooting
Performance test failure troubleshooting
4-
Table 4-9 Recommended adjustment for performance test failure
First Failed Test Recommended Adjustment
OCXO
Frequency Reference
AUX Input
Synthesizer Gain
Virtual Bridge Coefficient
Source Output Power
Receiver IF Range
Receiver Port Characteristics
Receiver Absolute Gain
Receiver RF Range
AUX Input ###
Frequency Accuracy (Standard) ###
Frequency Accuracy (1E5) ###
RF Output Level Accuracy &
Flatness
### ###
RF Output Level Linearity ### ###
Trace Noise
Crosstalk
System Dynamic Range
Noise Floor
Dynamic Accuracy ###
Uncorrected System
Performance
### ###
Table 4-10 Recommended adjustment for performance test failure
(Over 8.5 GHz Options)
First Failed Test Recommended Adjustment
OCXO
Frequency Reference
AUX Input
Synthesizer Gain
Source Output Power
Receiver IF Range
Receiver Port Characteristics
Receiver Absolute Gain
AUX Input ###