Keysight E5071C ENA Network Analyzers 273
Replacement Procedure
Inverter Board Replacement
Figure 6-40 Inverter Board Removal
Replacement Procedure
Step 1. Replace the inverter assembly by inverse procedure of removal.
Fasten the TORX T15 screws (item 2) using a torque screwdriver (set to
1.02 N-m / 9.0 lb-in).