Keysight E5071C ENA Network Analyzers 61
Adjustment
Required Adjustment after Replacing Assembly
Required equipment for the synthesizer gain adjustment
None
Virtual Bridge Coefficient Adjustment
This item writes the coefficient factors to the E5071C.
Required equipment for the virtual bridge coefficient adjustment
None
Source Output Power Adjustment
For up to 8.5 GHz Options, the purpose of this procedure is to adjust the
variable attenuator in Level Vernier board, and the gains among Level Vernier,
Distributor and Receiver.
For over 8.5 GHz Options, the purpose of this procedure is to adjust the
characteristic fluctuation of TC709C variable attenuator used in A86 Doubler
Board.
Required equipment for the source output power adjustment
1) Up to 8.5 GHz Options
2) Over 8.5 GHz Options
This item is for up to 8.5 GHz Options.
Description Recommended Model
Power Meter Keysight E4419A/B
Power Sensor Keysight E9304A Opt H18
Description Recommended Model
Power Meter Keysight E4419A/B
Power Sensor Keysight 8482A
Power Sensor Keysight E4413A
APC 3.5(m) - N(f)Adapter Keysight p/n 1250-1811
APC 3.5(f) -(m) Adapter Keysight p/n 85027-60005