60 Keysight E5071C ENA Network Analyzers
Adjustment
Required Adjustment after Replacing Assembly
3-
Required equipment for OCXO adjustment
Frequency Reference Adjustment
The purpose of this procedure is to adjust the 10 MHz frequency reference on
the source board.
Required equipment for frequency reference adjustment
AUX Input Adjustment
The purpose of this procedure is to adjust the offset error and proportional
error of AUX Input.
Required equipment for the AUX input adjustment
Synthesizer Gain Adjustment
The purpose of this procedure is to adjust 2Bit ATT of RF Path in Synthesizer for
Source.
Description Recommended Model
Frequency Counter Keysight 53181A with Opt. 010 and 124 or
Keysight 53131/2A with Opt. 010 and 124
Frequency Standard Symmetricom 5071A
BNC(m)-BNC(m) Cable, 61 cm Keysight p/n 8120-1839
Description Recommended Model
Frequency Counter Keysight 53181A with Opt. 010 and 124 or
Keysight 53131/2A with Opt. 010 and 124
Frequency Standard Symmetricom 5071A
BNC(m)-BNC(m) Cable, 61 cm Keysight p/n 8120-1839
Description Recommended Model
Function Generator Keysight 33120A or 33250A
Multi Meter Keysight 34401, 34410, 34411 or 3458A
BNC(m)-BNC(m) Cable, 61 cm x 4 ea Keysight p/n 8120-1839
BNC T Adapter x 2ea Common Parts
Banana BNC Common Parts