Keysight E5071C ENA Network Analyzers 47
Performance Test
6. NOISE FLOOR
6. NOISE FLOOR
Description
This test checks the noise floor for all the receiver ports of the ENA. The noise
floor is tested by performing load-ended absolute measurements with the test
ports terminated with "Load" devices, 16 times with segment sweep points for
specified frequency ranges. The measurement data is calculated from the RMS
deviation value from the 16 measurement data for each sweep frequency point.
Test equipment
Test Equipment (DSG)
1
1. The DSG (Device Substitution Group) is listed below the equipment description, if
applicable. These devices are required to be mapped in TME to execute a test.
DSGs require either GPIB communication or are important for dynamic measure-
ment uncertainty calculations. N/A refers to required equipment that is not
mapped as DSGs in TME.
Model Number Over 8.5
GHz
Option
2
2. Over 8.5 GHz are Options 2D5/2K5/4D5/4K5.
Up to
8.5 GHz
Option
3
3. Up to 8.5 GHz are Options
230/235/430/435/240/245/440/445/260/265/460/465/280/285/480/485.
Power Meter (Power Meter 1) N1914A X X
Power Sensor 1 (Power Sensor RF 1) E9340A (Opt H18) X
Power Sensor 2 (Power Sensor RF 2) N8482A X
Power Sensor 3 (Power Sensor MW 1) E4413A X
Load (2 or 4 required)
4
4. Two required for 2-port E5071C; four required for 4-port E5071C.
909D X
Load (2 or 4 required)*4 85032-60018 X
3.5 mm cable 8121-1834 X X
3.5 mm (f) to 3.5 mm (f) adapter (2 or
4 required)*4
85027-60005 X
Type-N (m) to 3.5 mm (f) (2 or 4
required)*4
1250-2879 X

Table of Contents