Keysight E5071C Service Manual -333
System Recovery
Calibration of the Touch Screen
Calibration of the Touch Screen
After E5071C measurement screen appears, you have to calibrate the touch
screen. Follow the procedure described below to calibrate the touch screen.
Step 1. Press .
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen.
The touch screen calibration screen (Figure B-26) appears.
Figure B-26 Touch Panel Calibration Screen
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also
on the upper right, lower left, and lower right. Touch the x marks in that order
with your finger.
Touching the four locations described above with your finger automatically
concludes the touch screen calibration.
e5071cse1078
With no operation on the touch screen calibration screen for a preset time,
it automatically closes and the previous measurement screen reappears.